Compare Gauss / Tesla Meters

                Model: 

 

MS175

IDR-325

Bell-5170

Bell-5270

Bell-5180

Bell-5280

 

Bell-6010 

Discontinued

Bell-8010

Bell-8030

Frequencies using  standard probes

 

DC-1kHz

DC-1kHz

DC-20 kHz

DC-25 kHz

 

DC-20 kHz

DC-50 kHz

DC-50 kHz

Probe included?

 

Yes

Yes

Yes

Yes

 

No

No

No

Triple-Axis or Single-Axis?

 

Single

Single

Single-axis

Single-axis

 

Single-axis

Single-axis

Triple-axis

Field Units Displayed

 

kA/m

 Gauss, kG

G, uT, or kA/m

G, uT,  or kA/m

 

G, uT, or kA/m

G, uT, or  kA/m

G, uT, or kA/m

Meter Size

 

Handheld

Handheld

Handheld

Handheld

 

Portable Benchtop

Portable Benchtop

Portable Benchtop

Communications Ports

 

None

RS-232

USB-C 2.0

USB-C 2.0

 

RS-232

USB, Ethernet, IEEE, RS232

USB, Ethernet, IEEE, RS232

LCD Display

 

3 digit

3 digit 

3½ digit

3½ digit

 

3½ digit

5 ¾ digit

5 ¾ digit

Rentals?

 

n.a.

Yes

n.a.

Yes

 

n.a.

n.a.

n.a.

 

 

 

 

 

 

 

 

 

 

Standard Probe:

 

Included

Included

Included

Included

 

Sold separately

Sold separately

Sold separately

    Max field

 

20 kA/m

(25 mT)

 30 kG

(3 T)

20 kG

(2 T)

30 kG

(3 T)

 

30 kG

(3 T)

30 kG

(3 T)

30 kG

(3 T)

    Resolution [a]

        (1X)

 

10 A/m

(12.6 uT)

 0.1 G

(10 uT)

0.1 G

(10 uT)

0.1 G

(10 uT)

 

0.1 G

(10 uT) 

0.1 mG

(0.01 uT)

0.1 mG

(0.01 uT)

    Accuracy [b]

        (1X)

 

±50 A/m

(±63 uT)

 ±0.4 G

(±40 uT)

±0.4 G

(±40 uT)

±0.4 G

(±40 uT)

 

±0.3 G

(±30 uT)

±3 mG

(±0.3 uT)

±3 mG

(±0.3 uT)

 

 

 

 

 

 

 

 

 

 

Low-field Probe: 

 

n.a.

n.a.

Sold separately

Sold separately

 

Sold separately

Sold separately

Sold separately

    Max field

 

 

 

1 G (100uT)

1 G (100uT)

 

2 G (200uT)

2 G (200uT)

2 G (200uT)

    Resolution [a]

        (0.01X)

 

 

 

1 mG

(0.1 uT)

1 mG

    (0.1 uT)

 

1 mG

(0.1 uT)

1 µG

(0.1 nT)

1 µG

(0.1 nT)

   Accuracy [b][c]

        (0.01X)

 

 

 

±4 mG

(±0.4 uT)

±4 mG

(±0.4 uT)

 

±3 mG

(±0.3 uT)

±0.03 mG

(±0.003 uT)

±0.03 mG

(±0.003 uT)

 

 

 

 

 

 

 

 

 

 

High-field Probe:

 

n.a.

n.a.

n.a.

n.a.

 

Sold separately

Sold separately

Sold separately

    Max field

 

 

 

 

 

 

30 to 300 kG  (3 to 30T)

30 to 300 kG (3 to 30T)

30 to 300 kG (3 to 30T)

    Resolution [a]

        (10X)

 

 

 

 

 

 

1 G

(100 uT)

1 mG

(0.1 uT)

1 mG

(0.1 uT)

    Accuracy [b]

        (10X)

 

 

 

 

 

 

±3 G

(±300 uT)

±30 mG

(±3 uT)

±30 mG

(±3 uT)

 

 

 

 

 

 

 

 

 

 

Notes:   n.a. means “not available”.    All meters use a Hall-effect type of probe. 

[a]  Resolution is Least Significant Digit. 1X, 0.01X, or 10X is resolution compared to standard probe, 0.01X is finer resolution.

[b] Accuracy shown is for measuring a weak DC (static) field. The ±error shown would be larger when measuring a strong field. AC field error may be larger than the DC error shown. For complete specifications see manual. 

 

[c] For best accuracy for Low-field probes: probes MOS51 and MOS81 are designed for use in a volume of uniform magnetic field around the probe at least 7000cc (0.25 cu.ft): for example a cylindrical volume 10cm (4") radius x 22cm (9") long, and MOS probes frequency response is DC to 700 Hz (-3dB point). 

Low-field probe MOX81 is designed for use in a volume of 84 liters (3 cu.ft) of uniform magnetic field, and MOX81 frequency response is DC to 400 Hz (-3db point). MOX81 also requires use of large zero-gauss chamber YA-112 sold separately. 

 

Conversion from A/m into Tesla or Gauss for measurements in air or free space:

         1 kA/m  = 1.26 mT  = 12.6 G
         1   A/m  = 1.26 uT   = 12.6 mG

The magnitude of the Earth's magnetic field at its surface ranges from 0.25 to 0.65 gauss, (0.025 to 0.065 mT), (250 to 650 mG), (25 to 65 μT).

For gaussmeters to measure AC magnetic fields only:  see AC Gaussmeter  tab at top of page.